Stratos LV/LV-T Technical Manual 151
11. Technical Data
11.1 Available Pacing Modes
DDDR, DDTRA, DDTR, DDIR, DDITR, DVIR, DVTR, DOOR,
VDDR, VDTR, VDIR, VVIR, VVTR, VOOR, AAIR, AATR, AOOR
DDD DDTA, DDT, DDI, DDIT, DVI, DVT, DOO, VDD, VDT, VDI,
VVI, VVT, VOO, AAI, AAT, AOO, OFF
VV synchronization for the Stratos CRT-Ps: BiV RV RV-T, OFF
11.2 Pulse- and Control Parameters
Basic Rate
32... (1)...60... (1)...88... (2)...122... (3)...140... (5)...180 ppm
Night Rate
Off, 32... (1)...88... (2)...122... (3)...140... (5)...180 ppm
Night Rate Start Time
00:00... (00:10)...23:50
Rate Hysteresis
Off; -5... (5)...-90 bpm
Repetitive Rate Hysteresis
Off; 1… (1)…15
Scan Rate Hysteresis
Off: 1… (1)…15
Upper Rate
90… (10)…180 ppm
UTR Response
2:1; Wenckebach (WKB)
Upper Tracking Rate, Atrium
Off, 200 ppm